TCH seal
TCH Laboratory — QMS Dashboard
Tarangnan Community Hospital · Quality Management System integrated with LUIB (read-only)
4
Notices
At a glance: 0 of 100 controlled documents approved (all V0 drafts) · 5 of 11 register indicators computable from this snapshot · visual library: 18 bench cards, 27 infographics, 29 flowcharts.
Snapshot freshness: Snapshot is 0 days old — within the weekly refresh cycle. (File: luib_snapshot.db, dated 2026-10-24 08:42)
Data source: this dashboard reads a snapshot — a copy of a LUIB backup file (see Snapshot freshness above), refreshed weekly. It is NOT the live clinical database and nothing here writes to LUIB. QMS document data is real: parsed from the actual MASTER-DOCUMENT-INDEX and the bundled document copies. All figures are computed live from the snapshot by this page.
Opened from LUIB? This dashboard runs in its own browser tab/window. To return to LUIB, close this tab or switch back to the LUIB window — nothing here writes to LUIB, and closing this page does not affect it.
Home / Documents / LAB-QM-ANA-006

LAB-QM-ANA-006 — Calibration and Verification of Ancillary Measuring Devices and Labware Policy

Type: Policy · Version: V0 · Status: DRAFT — UNAPPROVED · Served read-only from the bundled QMS copies. The Drive master is untouched.

Rendered in-browser from the project's Word file LAB-QM-ANA-006-calibration-and-verification-of-ancillary-measuring-devices-and-labware-policy.docx (policies-docx / forms-docx — the file Junry reviews). Formatting is faithful, not pixel-perfect; the DOCX is the official copy.

1.0 Purpose

The purpose of this policy is to establish defined procedures for the calibration, periodic verification, maintenance, and cleaning of ancillary measuring devices and laboratory glassware at Tarangnan Community Hospital Clinical Laboratory. This ensures volumetric accuracy, temperature precision, timing fidelity, and residue-free labware across all analytical testing procedures.

2.0 Scope

This policy applies to all mechanical pipettes, laboratory thermometers, digital timers, centrifuges, and reusable laboratory glassware across all sections of the Clinical Laboratory, governing all Medical Technologists and Laboratory Aides.

3.0 Definitions

4.0 Responsibility

4.1 Hospital Director

4.2 Pathologist

4.3 Chief Medical Technologist / Laboratory Quality Manager

4.4 Contracted Calibration Laboratory (External / Accredited Metrology Provider)

4.5 Medical Technologists

4.6 Laboratory Aides

5.0 Policy / Procedure

5.1 Policy Statement

The laboratory shall calibrate or verify all ancillary measuring devices affecting diagnostic examination quality before initial use and at scheduled intervals thereafter. Equipment calibration shall establish metrological traceability to certified national or international standards. Glassware shall be maintained clean, non-leaching, and chemically neutral to prevent sample contamination.

5.2 General Calibration and Verification Requirements

Device CategoryVerification FrequencyAcceptance CriteriaMethod / Reference Standard
Mechanical PipettesSemi-annuallyVolume delivery within ±1.0% to ±3.0% of nominalGravimetric water weight check (calibrated balance / multi-point)
In-Use ThermometersAnnually±1.0 °C of reference (refrigerators); ±0.5 °C (incubators)Comparison against NIST-traceable reference thermometer
Digital TimersSemi-annuallyWithin ±1.0 second per 10 minutesComparison against official NIST / atomic telephone reference time
CentrifugesAnnuallySpeed within ±2.0% of target RPM; timer within ±5.0%Photoelectric tachometer and calibrated stopwatch
Reusable GlasswarePer cleaning batchNeutral pH (6.5 to 7.5); no detergent residueBromothymol blue indicator rinse test

5.3 Pipette Calibration and Maintenance

5.4 Thermometer Calibration and Verification

5.5 Timer and Stopwatch Verification

5.6 Centrifuge Inspection and Speed Verification

5.7 Glassware Washing, Storage, and Neutrality Testing

5.8 Out-of-Tolerance Action and Remediation

6.0 Records

The following controlled record is maintained under this policy:

Form No.Form TitleParent Policy/SOPTypeStatus
Electronic recordAncillary Measuring Devices and Labware Calibration, Verification & Maintenance LogLAB-QM-ANA-006Log / RecordElectronic — registered under F-QM-002-B-V0

All calibration certificates, raw calibration data sheets, tachometer reports, and glassware pH logs shall be retained in accordance with Laboratory Retention Schedule Policy.

These records are maintained electronically in analyzer memory, the LIS, and dedicated electronic workbooks registered under F-QM-002-B-V0 (Decision D5); no separate paper form is maintained for them.

7.0 References

Companion policies: