TCH seal
TCH Laboratory — QMS Dashboard
Tarangnan Community Hospital · Quality Management System integrated with LUIB (read-only)
4
Notices
At a glance: 0 of 100 controlled documents approved (all V0 drafts) · 5 of 11 register indicators computable from this snapshot · visual library: 18 bench cards, 27 infographics, 29 flowcharts.
Snapshot freshness: Snapshot is 0 days old — within the weekly refresh cycle. (File: luib_snapshot.db, dated 2026-10-24 08:42)
Data source: this dashboard reads a snapshot — a copy of a LUIB backup file (see Snapshot freshness above), refreshed weekly. It is NOT the live clinical database and nothing here writes to LUIB. QMS document data is real: parsed from the actual MASTER-DOCUMENT-INDEX and the bundled document copies. All figures are computed live from the snapshot by this page.
Opened from LUIB? This dashboard runs in its own browser tab/window. To return to LUIB, close this tab or switch back to the LUIB window — nothing here writes to LUIB, and closing this page does not affect it.
Home / Documents / LAB-QM-ANA-005

LAB-QM-ANA-005 — Equipment Management, Calibration, and Temperature Monitoring Policy

Type: Policy · Version: V0 · Status: DRAFT — UNAPPROVED · Served read-only from the bundled QMS copies. The Drive master is untouched.

Related visual material

IG-23 — Equipment Tagout Return to Service

Rendered in-browser from the project's Word file LAB-QM-ANA-005-equipment-temperature-monitoring-policy.docx (policies-docx / forms-docx — the file Junry reviews). Formatting is faithful, not pixel-perfect; the DOCX is the official copy.

1.0 Purpose

The purpose of this policy is to establish an equipment management and environmental monitoring system for Tarangnan Community Hospital Clinical Laboratory. This governs equipment selection, installation qualification, inventory control, routine operation, preventive maintenance, calibration, downtime handling, decommissioning, and continuous temperature monitoring to guarantee uninterrupted, accurate diagnostic testing.

2.0 Scope

This policy applies to all analytical instrumentation, automated analyzers, centrifuges, microscopes, ancillary measuring devices, and temperature-dependent storage units (refrigerators, freezers, incubators) operated within the Clinical Laboratory, governing all Medical Technologists and Laboratory Aides.

3.0 Definitions

4.0 Responsibility

4.1 Hospital Director

4.2 Pathologist

4.3 Chief Medical Technologist / Laboratory Quality Manager

4.4 Servicing Equipment Distributor / External Provider (SAN PHARMA COMPANY)

4.5 Medical Technologists

4.6 Laboratory Aides

5.0 Policy / Procedure

5.1 Policy Statement

The laboratory shall manage all diagnostic equipment and temperature-dependent units through a structured life-cycle program comprising documented selection, qualification, maintenance, calibration, environmental monitoring, and decommissioning. Equipment shall be maintained in optimal operational condition, and temperature-dependent storage devices shall be monitored continuously to preserve specimen and reagent integrity.

5.2 Equipment Selection, Procurement, and Master Inventory

5.3 Installation Qualification and Commissioning

5.4 Equipment Identification and Usability Status

5.5 Operation and Authorized Personnel

5.6 Preventive and Routine Maintenance

5.7 Calibration, Calibration Verification, and Metrological Traceability

5.8 Equipment Temperature Monitoring and Acceptable Ranges

Equipment / AreaAcceptable Temperature RangeMonitoring Device
Reagent & Specimen Refrigerators2 °C to 8 °CCalibrated digital / spirit thermometer
Specimen & Control Freezers−15 °C to −25 °CCalibrated digital / probe thermometer
Incubators and Water Baths36.5 °C to 37.5 °CCalibrated immersion thermometer
Ambient Laboratory Workspace20 °C to 25 °C (Humidity 30%–70%)Calibrated thermo-hygrometer

5.9 Malfunction, Out-of-Service Protocols, and Decommissioning

Figure 1 — Equipment malfunction tag-out, repair, and return-to-service verification flow

5.10 Downtime Communication and Incident Escalation

6.0 Records

The following controlled record is maintained under this policy:

Form No.Form TitleParent Policy/SOPTypeStatus
F-QM-ANA-005-A-V0Master Equipment Inventory, Maintenance, Downtime & Temperature DossierLAB-QM-ANA-005Dossier / Master LogControlled

All equipment lifecycle records, service contracts, vendor PPM reports, and temperature logs shall be retained in accordance with Laboratory Retention Schedule Policy.