Power-On Sequence & Automated Startup PrimeSwitch on main power and computer terminal. Log into Pentra XL 80 software. Click Startup icon on toolbar; the analyzer automatically primes fluidics, zeroes optics, and runs background counts.SOP-H-002, Section 5.3.1
3
Verify Startup Background Blank Acceptance CountsVerify background counts meet strict limits: WBC < 0.3 × 10³/μL, RBC < 0.02 × 106/μL, HGB < 0.3 g/dL, PLT < 10 × 10³/μL. If high, repeat startup once or execute Backflush / Miniclean. Never test if blanks fail.SOP-H-002, Section 5.3.2, 5.3.3
4
Daily Tri-Level ABX Difftrol Quality Control RunEquilibrate Difftrol Low, Normal, and High vials at 18–25 °C for 15 minutes. Mix thoroughly by gentle rolling and inversion. Analyze in Quality Assurance > Control > Difftrol. Verify all parameters within ± 2 SD.SOP-H-002, Section 5.4.1
15 min eq
5
Patient Specimen Processing (Rack Loader / STAT Mode)Mix K2EDTA tubes on mechanical rotator ≥ 2 min + 8–10 inversions. Check for clots. Load 10-tube racks (barcodes outward) and press green Start, or aspirate pediatric / urgent tubes via manual STAT door.SOP-H-002, Section 5.5.1–5.5.3
6
End-of-Day Shutdown Enzymatic Cleaning (ABX Cleaner)Remove sample racks from loader deck. Clean exterior aspiration needles with alcohol wipe. Click Shutdown icon; analyzer draws ABX Cleaner to digest protein deposits and enters standby mode.SOP-H-002, Section 5.7.1
7
Weekly Concentrated Cleaning with ABX Minoclair (MANDATORY)Every Friday at shift close: Navigate to Services > Maintenance > Concentrated Cleaning. Aspirate 3.0 mL undiluted ABX Minoclair via manual port. Timed 15-min soak digests debris, followed by auto-flush.SOP-H-002, Section 5.8.1, 5.8.2