Tarangnan Community Hospital — Clinical Laboratory
Red Tag Out. Green Tag In.
Equipment malfunction — tag-out, repair, and verified return to service (LAB-QM-ANA-005 Figure 1).
INFOGRAPHIC IG-23
Source: LAB-QM-ANA-005
Version V0 — DRAFT
1 · Fault found — tag it out
🔴 RED TAG — OUT OF SERVICE
Defective performance, optical errors, or
repeated QC failures
? The instrument is
removed from service immediately
and tagged.
Review patient specimens
tested between the last satisfactory check and the failure.
The CMT
notifies clinical units
of testing delays.
Downtime and repair chronology are documented in the
equipment service log
.
Arrange alternative testing —
backup analyzer, referral laboratory, or STAT diversion
— per LAB-QM-IT-002.
2 · Prove it works again
After repair: perform
recalibration, precision checks, and routine QC testing
.
Decision:
does testing confirm satisfactory operational status?
3 · Return — or stay out
🟢 GREEN TAG — IN SERVICE
Satisfactory:
the CMT
authorizes return to service
; attach the Green IN SERVICE tag. Patient testing resumes.
Not satisfactory:
the instrument
remains Red-tagged OUT OF SERVICE
, quarantined from diagnostic use.
Fault / QC failures
→
RED tag — out of service
→
Review specimens + notify units
→
Repair + recalibrate + QC
→
CMT authorizes → GREEN tag