TCH seal
Tarangnan Community Hospital — Clinical Laboratory

POCT: Trained. QC'd. Verified.

Point-of-care testing — operator authorization, daily QC, and critical values (LAB-QM-POC-001 Figure 1).
INFOGRAPHIC IG-17
Source: LAB-QM-POC-001
Version V0 — DRAFT

1 · Who may test

2 · QC before patients

No — do not test: do not test patient samples. Tag the glucometer and quarantine it. Troubleshoot, re-QC, and verify return-to-service; notify the Chief Medical Technologist. A venous specimen is dispatched to the Clinical Laboratory for definitive confirmation.
Yes — in range: proceed to patient testing by skin puncture, applying capillary blood to the test strip, and document the measured glucose value in the patient medical chart.

3 · Critical or discordant result?

POCT records: the POCT Master List, device logs, and operator competency records are maintained electronically and registered under F-QM-002-B-V0 (LAB-QM-POC-001, Decision D5); no separate paper POCT form is maintained.
Train + certify operator
→
QC before patients
→
Out of range → tag & quarantine
→
Test + chart result
→
Critical? → venous verify + read-back