Tarangnan Community Hospital — Clinical Laboratory

Quality Indicators at a Glance

What we measure, the target for each, who owns it — trended monthly and quarterly by the CMT / LQM.
INFOGRAPHIC IG-02
Source: LAB-QM-003 +
F-QM-003-A-V0 · V0 DRAFT
IDIndicatorHow it is calculatedTargetFrequencyOwner
Pre-examination
QI-PRE-01Specimen Rejection RateRejected specimens ÷ total accessions received × 100< 1.0%MonthlyMedical Technologist (Receiving)
QI-PRE-02Request Form / Labeling CompletenessIncorrectly labeled or identified requests ÷ total requests × 100< 0.5%MonthlyMedical Technologist (Phlebotomy)
Examination
QI-ANA-01PT / EQAP Acceptable Performance RateSatisfactory PT results ÷ total PT challenges evaluated × 100100% (min 80%)Per survey / annualChief Medical Technologist
QI-ANA-02Internal QC Failure / Out-of-Control RateQC runs exceeding 13s or 22s limits ÷ total QC runs × 100< 2.0%MonthlyMedical Technologists (Bench)
QI-ANA-03*Equipment DowntimeTotal equipment out-of-service time recorded for the monthTrending — target to be setMonthlyMedical Technologists (Bench)
Post-examination
QI-POST-01Routine Test Turnaround Time (TAT)Routine tests meeting stated TAT ÷ total routine tests × 100≥ 95%MonthlyChief Medical Technologist
QI-POST-02STAT Test TAT ComplianceSTAT tests released within 60 min ÷ total STAT tests × 100≥ 98%MonthlyChief Medical Technologist
QI-POST-03Critical Values Notification ComplianceCritical values called within 15 min ÷ total critical values × 100100%MonthlyMedical Technologists (Testing)
QI-POST-04Corrected / Amended Report RateCorrected / amended reports ÷ total reports issued × 100< 0.2%MonthlyChief Medical Technologist
General QMS
QI-GEN-01Customer & Patient Satisfaction RatePositive satisfaction ratings ÷ total survey responses × 100≥ 90%Quarterly / annualChief Medical Technologist
QI-GEN-02*Monthly Test Volume & WorkloadTotal tests performed in the month (count)Trending only — no targetMonthlyChief Medical Technologist
TARGET MISSED
OR BAD TREND?
1 · OccurrenceAcute incident → LAB-QM-010
→
2 · CAPASystemic / recurring → LAB-QM-011
→
3 · Risk updateNew hazard → LAB-QM-009
→
4 · CQI studyBroader improvement → F-QM-003-B-V0
Note: * In the LAB-QM-003 §5.3 core list but previously had no register entry. Added by Junry's decision, 2026-10-21 — IDs/owners follow the register's pattern, to be confirmed when F-QM-003-A-V0 is updated. Results go to the CQI Committee and Management Review (§5.4).